RMM Device Pairing and Measurement Guide

  • For pairing medical devices to the RMM app and how to receive measurements.


    Bodimetrics O2 Vibe

    Pairing

    1. Press the button to power the device on
    2. The mobile app should identify the device after no more than a few seconds
    3. The mobile app should complete its query of the device after no more than a few seconds
    4. After pairing is complete, hold the button to power the device off

     Measurement

    1. Put the device around your wrist and attach the finger cuff
    2. Press the button to power the device on
    3. The device will start transmitting previously captured measurements that had not been exported
    4. The device will then send current measurements periodically as long as you are wearing the device
    5. Press and hold the button to power the device off

    Bodimetrics GBF1246B/GBF1502B

    Pairing

    1. Press the pairing button on the bottom side of the device
    2. The mobile app should identify the device after no more than a few seconds
    3. The mobile app should complete its query of the device after no more than a few seconds
    4. After pairing is complete, exit the device’s bluetooth mode by pressing the home button

     Measurement

    1. Press the SET button to get to the User Select Screen
    2. If you have not set up a User Profile
      1. Select one of the User’s (P1,P2,P3, etc)
      2. Follow the prompts to enter your Age and Height
    3. Step on the scale, making sure your bare feet are making contact with all four electrodes

    Bodimetrics BM 88

    Pairing

    1. Power the device on, tap the right arrow to get to the second page, and press the bluetooth icon.
    2. The mobile app should identify the device after no more than a few seconds
    3. The mobile app should complete its query of the device after no more than a few seconds
    4. After pairing is complete, let the device power itself off, this may take a minute

     Measurement

    1. Choose a device mode you would like to take a measurement with
    2. Choose a user profile on the device.
      1. Be sure the User ID# number matches the Patient ID from the app
    3. Take the measurement
    4. Press the export button (rightmost button) to enable bluetooth transmission
    5. Once the measurements have shown up on the Measurement Page on the app, press the home button on the device.

    Bodimetrics Wrist BP Cuff

    Pairing

    1. Power the device on, you do not have to be wearing it.
    2. The mobile app should identify the device after no more than a few seconds
    3. The mobile app should complete its query of the device after no more than a few seconds
    4. After pairing is complete, power the device down

     Measurement

    1. Attach the Cuff to your wrist
    2. Power on the device and wait for it to complete its measurement
    3. As soon as the final result is displayed, the measurement will be sent to the app
    4. Power down the device after the measurement has been transferred

    ForaCare TNG Voice 4282

    Pairing

    1.  Slide the button on the right of the device up, hold for 5 seconds, then release
      1. After 1-2 seconds you should see the LED in the upper left corner flash blue
    2. The mobile app should identify the device after no more than a few seconds
    3. The mobile app should complete its query of the device after no more than a few seconds
      1. When the query is in progress, the flashing blue light should remain solid
    4. After pairing is complete, the device must be shut of manually by pressing the front button until it reads OFF

     Measurement

    1. The device will send the measurement as soon as it has been taken.
    2. The LED in the upper left will start flashing blue, then will become solid once the mobile app has made the connection
    3. Once the measurement has been captured, the mobile app will turn off the device automatically

    ForaCare TNG BP 3140

    Pairing

    1. Press the smaller of the two buttons, on the right hand side of the device.
      1. After 1-2 seconds the screen will read “PCL” and you should see the LED on the left flash blue
    2. The mobile app should identify the device after no more than a few seconds
    3. The mobile app should complete its query of the device after no more than a few seconds
      1. When the query is in progress, the flashing blue light should remain solid
    4. After pairing is complete, the device must be shut of manually by pressing the larger of the two buttons, on the left side, at which time the screen will turn off

     Measurement

    1.  The device will bond with the mobile app as soon as the measurement process begins
      1. After turning the device on, the LED will flash blue, then remain solid once connected and throughout the measurement process
    2. When the measurement is complete, it will be sent to the mobile app
    3. Once the measurement has been captured, the mobile app will turn off the device automatically

    ForaCare D40 3261

    Pairing

    1. Press the Bluetooth button, on the front/right of the device.
      1. After 1-2 seconds the screen will read “PCL” and you should see the LED on the left flash blue
    2. The mobile app should identify the device after no more than a few seconds
    3. The mobile app should complete its query of the device after no more than a few seconds
      1. When the query is in progress, the flashing blue light should remain solid
    4. After pairing is complete, the device must be shut off manually by pressing the Power button, on the front/right, at which time the screen will read OFF

     Measurement

    1. The device will send the measurement as soon as it has been taken.
    2. The LED in the upper left will start flashing blue, then will become solid once the mobile app has made the connection
    3. Once the measurement has been captured, the mobile app will turn off the device automatically

    The behavior is the same for both blood pressure and glucose measurements


    ForaCare TNG Scale 550

    Pairing

    1. Press and hold SET for 5 seconds until the bluetooth symbol starts flashing on the device
    2. The mobile app should identify the device after no more than a few seconds
    3. The mobile app should complete its query of the device after no more than a few seconds
    4. After pairing is complete, the device must be shut of manually by pressing the Up or Down arrow buttons.

     Measurement

    1. The device will send the measurement as soon as it has been taken.
    2. The bluetooth symbol in the upper right will start flashing blue, then will become solid once the mobile app has made the connection
    3. Once the measurement has been captured, the mobile app will turn off the device automatically

    ForaCare Thermometer IR20b

    Pairing

    1. Take a reading with the thermometer to start broadcasting the bluetooth from the device

      1. A blue light on the bottom of the device should start flashing blue

    2. The mobile app should identify the device after no more than a few seconds
    3. The mobile app should complete its query of the device after no more than a few seconds
      1. The blue light on the bottom of the device should be solid blue
    4. After pairing is complete, the mobile app will turn off the device automatically

    The initial reading taken in Step 1 is only used to initialize bluetooth paring and is NOT recorded and pushed online

    Measurement

    1. The device will send the measurement as soon as it has been taken
    2. The blue light on the bottom of the device will start flashing blue, then will become solid once the mobile app has made the connection
    3. Once the measurement has been captured, the mobile app will turn off the device automatically

    Nonin 3230

    Pairing

    1. Place the Device on your finger
      1. A splash screen will display immediately
      2. After 3-5 seconds you should see the measurement screen with a bluetooth icon in the upper right
    2. The mobile app should identify the device after no more than a few seconds of this screen being displayed
    3. The mobile app will then attempt to complete its query of the device
      1. When the query is in progress, the bluetooth icon in the upper right will turn green
        1. On the older 3230 devices, this process could take 15-30 seconds
        2. On the v2.7 or newer 3230 devices, this process will only take 1-2 seconds
    4.  After pairing is complete, the device must be shut off, by removing it from the finger and letting it turn off
      1. After 5-10 seconds, the screen will turn off
        1. On the older 3230 devices, it will take another 20-30 seconds to complete the disconnection process
        2. On the v2.7 or newer 3230 devices, this process will only take 1-2 seconds

     Measurement

    The mobile application will only capture every 10th measurement by default, so you will only see a new measurement in the app every 10 seconds

    This interval can be configured in the settings,

    1. The device will send a measurement once per second
    2. The bluetooth icon will turn green once the connection to the mobile app has been established
    3. Measurements will be captured as long as the device is active and in range.
    4. Remove the device from your finger to stop sending measurements.
      1. After 5-10 seconds, the screen will turn off
        1. On the older 3230 devices, it will take another 20-30 seconds to complete the disconnection process
        2. On the v2.7 or newer 3230 devices, this process will only take 1-2 seconds

    Omron HEM-9200T

    Pairing

    1. Take a reading with the device
    2. When the result is displayed, the mobile app can then detect the device
    3. The app will ask for a PIN, this is located on the bottom of the device, by the Manufacturer Device information
    4. After the device has been paired, turn the device off

     Measurement

    1. The device will send a measurement as soon as it has been taken
    2. The parallel arrows icon will be lit while the device is communicating with the app
    3. After the measurement has been transferred, power off the device